The diagnostic potential of MicroDTTect compared to conventional culture of tissue samples in orthopedic infections

Department of life, Health & Environmental Sciences, Unit of Orthopaedics and Traumatology, University of L'Aquila, L'Aquila, Italy
https://orcid.org/0000-0002-7511-1288
Department of life, Health & Environmental Sciences, Unit of Orthopaedics and Traumatology, University of L'Aquila, L'Aquila, Italy
Department of life, Health & Environmental Sciences, Unit of Orthopaedics and Traumatology, University of L'Aquila, L'Aquila, Italy
https://orcid.org/0000-0001-7766-0806
Unit of Orthopaedics and Traumatology, San Liberatore Hospital, Atri (TE), Italy
Laboratory of Clinical Pathology, San Liberatore Hospital, Atri (TE), Italy
Unit of Orthopaedics and Traumatology, San Liberatore Hospital, Atri (TE), Italy
Department of life, Health & Environmental Sciences, Unit of Orthopaedics and Traumatology, University of L'Aquila, L'Aquila, Italy
orthopedic infections periprosthetic infections MicroDTTect

Abstract

Objective. The evolution of new prosthetic and osteosynthesis devices has led to better outcomes and therefore more frequent surgical indications. As a consequence, an increased incidence of complications such as infections or aseptic loosening in orthopedic and trauma surgery are being recorded in general population. MicroDTTect is a quick and simple system that is useful to detect low grade or bio-film related infections. The aim of this study was to evaluate the reliability of MicroDTTect in detection of orthopedic infections compared to conventional culture of tissue samples.
Methods. The population enrolled was composed of 13 patients undergoing surgery for prosthesis or osteosynthetic device failure or loosening. The MicroDTTect system and traditional culture of tissue samples were applied to identify the pathogens and compared with each other.
Results. MicroDTTect had a higher sensitivity compared to conventional culture of tissue samples. Two cases resulted positive while the traditional culture sample showed a false negative result. In addition, with MicroDTTect a polymicrobial infection was identified, while with traditional methods was misdiagnosed.
Conclusions. We showed that treatment of suspected implant infections using the MicroDTTect device improves microbiological diagnosis with more sensitive results, leading to a more accurate treatment.

Introduction

The evolution of new prosthetic and osteosynthesis devices has led to better outcomes and therefore increasingly frequent surgical indications. This is accompanied by an increased incidence of complications related to the implants and to the surgical intervention.

Among complications, chronic post-operative bacterial infections play an important role associated with primary implants (1-4%) and revision surgery (30%) 1-4.

Chronic infections due to low-virulence organisms can be difficult to diagnose and challenging to treat, leading to high morbidity rates and high healthcare costs 5. Clinical symptoms, inflammatory markers, and scintigraphy may help in identifying chronic infections, but the diagnostic benchmark is considered to be intra-operative clinical diagnosis and pathogen identification, in order to administer tailored antibiotic treatment 6-8.

It should be taken into consideration that implant-related infections are biofilm-related 9,10 and traditional sampling techniques may not be so reliable to detect biofilm-embedded bacteria from implant surfaces and peri-implant tissues, leading to possible false negative results in culture, and to a misdiagnosis of aseptic loosening, in particular when recent antibiotic treatment or low-grade infections place doubt on clinical signs. Sonication of the explanted implants and culture of the sonication fluids has been considered superior over other methods to disrupt the bacterial biofilm 10-14.

This procedure is, however, not used in all the laboratories given its purchase cost and difficulty of use.

Some authors proposed an alternative technique using dithiothreitol (DTT) to detach biofilm from explanted implants with good results 15,16. DTT, in fact, is a chemical agent that reduces disulfide bonds in peptides, and can thus alter the matrix of biofilm releasing bacteria without affecting their viability. Accordingly, bacterial culture is possible and permits identification and antibiotic susceptibility tests.

Recently, a commercial device containing a DTT solution has been developed, namely MicroDTTect (4i for infection, Monza, Italy) (Fig. 1). This device is a sterile closed system that avoids multiple manipulation decreasing the risk of contaminations, and is quick and easy to use.

The purpose of the study was to evaluate the reliability of MicroDTTect in prosthetic and implant failures in detecting chronic or silent infections compared to the conventional culture of tissue samples.

Materials and methods

A population of 13 patients undergoing surgery for suspected septic complications or aseptic failure was enrolled. Nine patients underwent prosthesis revision, 3 patient antibiotic cement spacer removal, and 1 plate removal. The demographic and implant characteristics are detailed in Table I.

Both the MicroDTTect system and the collection method used in our institute (culture of tissue samples) were applied to identify the pathogens in the study group. We evaluated the number of positive and negative samples to compare MicroDTTect methodology with tissue sample culture.

Prosthetic implants were aseptically collected and immediately placed in the MicroDTTect collection device (Fig. 2) avoiding unnecessary manipulations (Fig. 3). In parallel, periprosthetic tissue samples (PPT) were collected and immediately placed in plastic sterile containers. Samples were transported as soon as possible to the laboratory for microbiological analysis.

The MicroDTTect procedure was performed according to the manufacturer’s instructions. The MicroDTTect device was placed on a shaker for 15 min. to allow the DTT to detach bacteria and biofilm from the prosthetic material. The DTT suspension was aseptically collected, centrifuged for 10 min. at 3000 x g and the supernatant except for 1 mL was discarded. The pellet was resuspended in the remaining DTT solution. 100 uL aliquots of DTT-treated sample were seeded onto sheep blood agar (COS), McConkey agar (MCK), mannitol salt agar (MSA), Sabouraud agar (SGC), chocolate agar (PVX), Schaedler agar (SCS) (bioMerieux, Marcy l’Etolie, France), Bi-state blood culture bottle (BCB) (Autobio diagnostics, Zhengzhou, China) and thioglycollate broth (TB) (Liofilchem, Roseto, Italy). PPT were cultured in BCB and TB. COS, MCK, MSA and SGC were incubated for 48 h at 37°C; PVX was incubated for 48 h at 37°C in 5% CO2 enriched atmosphere; SCS was incubated for 48 h in anaerobic conditions; BCB and TH were incubated for 7 days at 37°C and observed daily: positive broths were sub-cultured on agar plates (COS, MCK, MSA, SGC, PVX, SCS) for 48 h.When cultures were positive, identification was performed at the phenotypic level by Vitek2 Compact (bioMerieux, Marcy l’Etolie, France).

Results

MicroDTTect had a higher sensitivity compared to conventional culture of tissue samples. In fact, bacteriological growth was observed in 5 of 13 (38.5%) implant samples using the MicroDTTect system, while PPT samples were positive by culture for only 3 of 13 patients (23%). Two implants resulted positive with MicroDTTect device and negative with traditional culture of tissue samples, respectively, for Escherichia coli and Staphylococcus hominis ssp Hominis. Furthermore, one implant resulted positive for 2 microorganisms, Proteus mirabilis and Enterococcus faecium, while positive for only Proteus mirabilis in traditional tissue samples (Tab. II).

Agreement between the two procedures was found in 10 of 13 patients (77%), 8 of them resulting sterile, 2 resulting infected from the same bacteria (one implant was infected from Serratia marcenses and one implant from Staphylococcus hominis ssp Hominis and Staphylococcus lugdunensis).

There was no microorganism identified only in tissue samples.

Discussion

The quality of the implants used in orthopedics and trauma surgery and the surgical techniques and approach are improving clinical results and satisfaction of patients and surgeons. As a result, an increasing number of implant surgeries are being performed. Unfortunately, the increase in the surgical management is inevitably accompanied by an increasing incidence of peri-implant infections 1,2.

Diagnosis of orthopedic infections can be challenging and is always based on a combination of clinical, biological, and microbiological findings. However, when a chronic infection has to be treated, the clinical signs and laboratory results might be dubious, especially when recent empiric antibiotic therapy has been administered leading to false negative results.

Biofilm formation in chronic infections is considered as a major cause of the insufficient sensitivity of classical culture approaches using tissue samples. Sonication has shown to notably increase the sensitivity of microorganism identification by detachment of bacteria biofilm covering the orthopedic implant, and many Authors documented the superiority of sonication in comparison with tissue culture, with a lower sensitivity for the latter one (ranging from 61 to 76%) with respect to sonicated implants (77-95%) 17.

Previous studies 13,14,16 have shown that a chemical agent, namely DTT, is a reliable alternative to sonication for microbiological diagnosis of orthopedic infections and may be even more sensitive than sonication towards S. epidermidis, which is often involved in peri-implant infections.

As sonication devices are not available in all laboratories, it was decided to undertake this study with the aim of testing the efficacy and reliability of a commercial device that takes advantage of the properties of DTT to disrupt bacterial biofilm, called MicroDTTect.

Other authors have investigated the properties of MicroDTTect in daily clinical practice. Calori et al. 16, for example, reported a higher sensitivity of MicroDTTect in analyzing both prosthetic and osteosynthetic devices compared to a control group. However, the control group, was composed of samples collected with flocked swabs, which are not considered a suggested method for periprosthetic infections 18,19.

Kolenda et al. 19 described a higher sensitivity of MicroDTTect compared with tissue sample cultures, but in their study only prosthetic joint infections were investigated.

In our study, periprosthetic, periosteosynthetic device, and pericement spacer infections were investigated when there was failure and loosening of the implant. As sustained by Kolenda et al. 20 and Sambri et al. 14 in their studies, low-grade and chronic infections can be difficult to detect and DTT treatment and MicroDTTect may be useful especially in cases when peri-implant infection is not suspected preoperatively.

Although with a low level of evidence, our data confirms the results of Calori 16 and Kolenda 20. MicroDTTect had greater sensitivity compared to conventional culture of tissue samples. In two cases, in fact, there was a positive result (for Escherichia coli and Staphylococcus hominis ssp Hominis respectively) with MicroDTTect, while negative with traditional culture of tissue sample. In one patient, MicroDTTect allowed us to detect a polymicrobial infection due to Proteus mirabilis and Enterococcus faecium, while only Proteus mirabilis was isolated with conventional tissue culture (Tab. II).

MicroDTTect is a quick and easy device to use, with minimal manipulation needed and with a closed system that guarantees safe and sterile transportation of the material to the laboratory, thus reducing the risk of bacterial contamination.

In addition, Romanò et al. 17 also suggested that the use of the MicroDTTect device may be cost-effective. In fact, although the immediate direct costs are increased, the extra costs generated by diagnostic inaccuracy of traditional tissue culture may increase indirect costs due to time required for sample treatment, useless or unnecessary medical treatments, and possible medical claims.

As the Author reported in his manuscript, for every 100 patients treated each year the total cost of a wrong diagnosis with a tissue culture would amount to € 1.6 mln; with sonication the total potential costs would be reduced to respectively € 1.4 mln and with MicroDTTect to € 0.7 mln 17.

The limitations of the study are the small number of the cases analyzed. Thus, studies on a larger population are needed as well as a clinical trial to validate the use of the device.

Conclusions

In conclusion, the use of MicroDTTect device in implant loosening management improved microbiological diagnosis with more sensitive results by allowing identification of additional bacteria compared to traditional culture.

High morbidity rates and high healthcare costs are related to chronic infections, and more accurate and quicker identification of the pathogen may increase the appropriateness and efficacy of antibiotic therapy. This could and reduce the severity of sequelae, recovery time, and costs of treatment while improving patient’s quality of life.

Acknowledgements

We thank the laboratory technicians Candido Ruggieri, Roberta Caporaletti and Antonia Ranieri for the contribution and the excellent work done.

Conflict of interest statement

None of the Authors has any financial and personal relationships with other people or organizations that might pose a conflict of interest in connection with the submitted article.

Funding

This research did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors.

Authors’ contributions

AG, JR, AF, MDM: have made substantial contributions to the conception and design of the work; AG, JR: have been involved in acquisition of data and drafting the manuscript; MB, FF: have contributed with analysis and interpretation of data; FF, VC: revised the manuscript critically and have given final approval of the version to be published. All Authors read and approved the final manuscript.

Ethical consideration

All patients were treated according to the ethical standards of the World Medical Association’s Declaration of Helsinki and were invited to read, understand and sign the dedicated informed consent form.

Figures and tables

Figure 1.MicroDTTect device.

Figure 2.Explanted prosthesis in the MicroDTTect bag.

Figure 3.Avoiding manipulation of the explanted implant.

Males (%) 5 (38.5%)
Females (%) 8 (61.5%)
Age in years, mean ± SD 69 ± 5
Total hip arthroplasty (%) 5 (38.5%)
Total knee arthroplasty (%) 3(23%)
Total shoulder arthroplasty (%) 1 (7.7)
Hip spacer (%) 1 (7.7%)
Knee spacer (%) 1 (7.7%)
Shoulder spacer (%) 1 (7.7%)
Ulnar plate (%) 1 (7.7%)
Table I.Population characteristics.
Removed implant Tissue sample results MicroDTTect results
1 Total hip arthroplasty (THA) Sterile Sterile
2 THA Sterile Sterile
3 THA Staphylococcus hominis ssp Hominis, Staphylococcus lugdunensis Staphylococcus hominis ssp Hominis, Staphylococcus lugdunensis
4 Knee antibiotic spacer Proteus mirabilis Proteus mirabilis, Enterococcus faecium
5 Total knee arthroplasty (TKA) Sterile Sterile
6 Ulnar plate Sterile Sterile
7 TKA Sterile Sterile
8 THA Sterile Sterile
9 THA Sterile Escherichia coli
10 TKA Serratia marcenses Serratia marcenses
11 Hip antibiotic spacer Sterile Sterile
12 Shoulder antibiotic spacer Sterile Sterile
13 Total shoulder arthroplasty (TSA) Sterile Staphylococcus hominis ssp Hominis
Table II.Isolation of microorganisms with MicroDTTect and tissue samples.

References

  1. Urquhart DM, Hanna FS, Brennan SL. Incidence and risk factors for deep surgical site infection after primary total hip arthroplasty: a systematic review. J Arthroplasty. 2010; 25:1216-1222. DOI
  2. Bozic K, Kurtz SM, Lau E. The epidemiology of revision total hip arthroplasty in the United States. J Bone Joint Surg Am. 2009; 91:128-133. DOI
  3. Antoci V, Adams CS, Hickok NJ. Vancomycin bound to Ti rods reduces periprosthetic infection: preliminary study. Clin Orthop Relat Res. 2007; 461:88-95. DOI
  4. Giavaresi G, Borsari V, Fini M. Preliminary investigations on a new gentamicin and vancomycin-coated PMMA nail for the treatment of bone and intramedullary infections: an experimental study in the rabbit. J Orthop Res. 2008; 26:785-792. DOI
  5. Laurent E, Gras G, Druon J. Key features of bone and joint infections following the implementation of reference centers in France. Médecine Mal Infect. 2018; 48:256-262. DOI
  6. Trampuz A, Piper KE, Jacobson MJ. Sonication of removed hip and knee prostheses for diagnosis of infection. N Engl J Med. 2007; 357:654-663. DOI
  7. Piper KE, Jacobson MJ, Cofield RH. Microbiologic diagnosis of prosthetic shoulder infection by use of implant sonication. J Clin Microbiol. 2009; 47:1878-1884. DOI
  8. Della Valle C, Parvizi J, Bauer TW. The diagnosis of periprosthetic joint infections in the hip or knee. Guideline and evidence report. Journal American Academy of Orthopedic Surgeons. 2010. DOI
  9. Parvizi J, Zmistowski B, Berbari EF. New definition for periprosthetic joint infection: from the workgroup of the musculoskeletal infection society. Clin Orthop Relat Res. 2011; 469:2992-2994. DOI
  10. Holinka J, Bauer L, Hirschl AM. Sonication cultures of explanted components as an add-on test to routinely conducted microbiological diagnostics improve pathogen detection. J Orthop Res. 2011; 29:617-622. DOI
  11. Dudareva M, Barrett L, Figtree M. Sonication versus tissue sampling for diagnosis of prosthetic joint and other orthopedic device-related infections. J Clin Microbiol. 2018; 56:e00688-18. DOI
  12. Drago L, Romanò CL, Mattina R. Does dithiothreitol improve bacterial detection from infected prostheses? A pilot study. Clin Orthop Relat Res. 2012; 470:2915-2925. DOI
  13. Drago L, Signori V, De Vecchi E. Use of dithiothreitol to improve the diagnosis of prosthec joint infections. J Orthop Res. 2013; 31:1694-1699. DOI
  14. Sambri A, Cadossi M, Giannini S. Is treatment with dithiothreitol more effective than sonication for the diagnosis of prosthetic joint infection?. Clin Orthop. 2018; 476:137-145. DOI
  15. De Vecchi E, Bortolin M, Signori V. Treatment with dithiothreitol improves bacterial recovery from tissue samples in osteoarticular and joint infections. J Arthroplasty. 2016; 31:2867-2870. DOI
  16. Calori GM, Colombo M, Navone P. Comparative evaluation of MicroDTTect device and flocked swabs in the diagnosis of prosthetic and orthopaedic infections. Injury. 2016; 47:S17. DOI
  17. Romanò CL, Trentinaglia MT, De Vecchi E. Cost-benefit analysis of antibiofilm microbiological techniques for peri-prosthetic joint infection diagnosis. BMC Infect Dis. 2018; 18:154. DOI
  18. Osmon DR, Berbari EF, Berendt AR. Diagnosis and management of prosthetic joint infection: clinical practice guidelines by the Infectious Diseases Society of America. Clin Infect Dis Off Publ Infect Dis Soc Am. 2013; 56:e1-e25. DOI
  19. Gehrke T, Parvizi J. Proceedings on the international consensus meeting on periprosthetic joint infection. Bone Joint J. 2018; 95-B:1450-1452. DOI
  20. Kolenda C, Josse J, Batailler C. Experience with the use of the MicroDTTect device for the diagnosis of low-grade chronic prosthetic joint infections in a routine setting. Front Med (Lausanne). 2021; 8:565555. DOI

Affiliations

Alessio Giannetti

Department of life, Health & Environmental Sciences, Unit of Orthopaedics and Traumatology, University of L'Aquila, L'Aquila, Italy

Jacopo Romano

Department of life, Health & Environmental Sciences, Unit of Orthopaedics and Traumatology, University of L'Aquila, L'Aquila, Italy

Andrea Fidanza

Department of life, Health & Environmental Sciences, Unit of Orthopaedics and Traumatology, University of L'Aquila, L'Aquila, Italy

Michele Di Mauro

Unit of Orthopaedics and Traumatology, San Liberatore Hospital, Atri (TE), Italy

Mauro Brunetti

Laboratory of Clinical Pathology, San Liberatore Hospital, Atri (TE), Italy

Fabrizio Fascione

Unit of Orthopaedics and Traumatology, San Liberatore Hospital, Atri (TE), Italy

Vittorio Calvisi

Department of life, Health & Environmental Sciences, Unit of Orthopaedics and Traumatology, University of L'Aquila, L'Aquila, Italy

Copyright

© © Ortopedici Traumatologi Ospedalieri d’Italia (O.T.O.D.i.) , 2022

How to Cite

[1]
Giannetti, A., Romano, J., Fidanza, A., Di Mauro, M., Brunetti, M., Fascione, F. and Calvisi, V. 2022. The diagnostic potential of MicroDTTect compared to conventional culture of tissue samples in orthopedic infections. Lo Scalpello - Journal. 36, 2 (Sep. 2022), 111-115. DOI:https://doi.org/10.36149/0390-5276-262.
  • Abstract viewed - 540 times
  • PDF downloaded - 181 times